The KT 200 Kjeltec™ Distillation unit provides a simple and reliable solution for safe and semi-automatic distillation for protein determination. The possibility to pre-program alkali addition and distillation time supports the production of accurate results independent of operator.
Simple, safe and reliable analysis according to official standards for accurate results independent of operator.
Food, feed, feed ingredients, forage, pet food, grain, cereal, oil seeds, meat, meat products, milk, dairy products, beer and beer ingredients.
Nitrogen, protein, Non Protein Nitrogen (NPN), and volatiles (SO2, TVBN and more).
A high accuracy bellows pump gives long term stability in delivery of the alkali. The unit is equipped with a wide range of safety features including safety door, plus level pins and overpressure sensors for the steam generator. The door completely covers the area in front of the operator and the unit cannot be operated with the door open. The drip tray can easily be removed for cleaning.
Official and accurate procedures (ISO, AOAC, EPA, DIN) simplify validation.
The patented SAfE* technology improves safety by a mixing procedure, reducing the exothermic reaction between alkali and acid during distillation and removing the need for manual pre-dilution. *Steam Addition for Equilibration
| Performance data at 230V | |
| Analysis time | 4 minutes at 30 mg N (7 minutes at 200 mg N) |
| Distillation capacity | ~ 30 ml/min |
| Measuring range | 0.1 – 200 mg N |
| Reproducibility | 1% RSD (including the digestion step) |
| Recovery |
> 99.5% at nitrogen levels between 1 – 200 mg N |
| Installation Requirements |
|
| Power supply | 200 – 240 V 50 – 60 Hz |
| Power consumption | 2 200 W |
| Water consumption (during distillation only) |
2 l/min at water temp of 30°C (During distillation only) 1 l/min at water temp of 15°C |
| Weight net | 25 kg |
| Dimensions (W x D x H) | 300 × 280 × 770 mm |
The Kjeltec™ series consists of four models: Kjeltec™ 8100, 8200 and 8400 and KT 200 Kjeltec™ with different levels of automation.
| 200 | 8100 | 8200 | 8400 | |
| Variable output steam generator | x | x | x | |
| Water saving system | x | x | x | |
| Cooling water flow control | x | x | x | |
| Distillate temperature monitor | x | x | x | |
| Polypropylene splash head | x | x | x | x |
| Polypropylene tube emptying vessel | x | x | x | |
| Bellows pumps for reagent addition | x | x | x | x |
| Automatic alkali addition | x | x | x | x |
| Automatic dilution water addition |
|
x | x | x |
| Automatic receiver solution addition | x | x | ||
| SAfE - Steam Addition for Equilibration | x | x | x | x |
| Tube emptying/Waste collection | x | x | x | |
| Reagent alarms | x | x | x | |
| Removable drip tray | x | x | x | x |
| Automatic safety door | x | x | ||
| Interlocked safety door | x | x | x | |
| Removable safety door | x | x | x | |
| Tube-in-place safety | x | x | x | |
| Tube replacement sensor | x | x | x | |
| Colorimetric titration system | Opt. | x | ||
| External titrator connection | Opt. | Opt. | ||
| Interchangeable burette | x | |||
| Titration bypass | x | |||
| Results calculation | x | |||
| Storage of raw data in instrument | 40 batch | |||
| Data management PC software | x | |||
| Storage of raw data with PC software | Unltd | |||
| Ethernet connection for PC | x | |||
| Balance | x | |||
| Printer | x | |||
| Autosampler system 8420 or 8460 | Opt. |