MycoFoss™

MycoFoss™

MycoFoss™ is the first ever mycotoxin testing solution to combine automation and speed in one. Just place the sample and press start. Results are delivered within eight minutes.

Value

Value

Make fast decisions on raw material intake, segregation, process adjustments and the use of additive binders and ensure compliance with legal requirements.

Type sample

Type sample

Maize, wheat and barley

Parameters

Parameters

MAIZE AND WHEAT: Aflatoxins, deoxynivalenol, fumonisins, zearalenone, ochratoxin A, trichothecene 2 | BARLEY: deoxynivalenol

Smarter management of mycotoxin risks in the grain industry

Smarter management of mycotoxin risks in the grain industry

For grain producers and flour millers, protecting consumers from the harmful effects of mycotoxin contamination is of utmost importance. Learn how fast, accurate and reliable results make it easier for producers to check contamination levels and match international food safety requirements.
Smarter management of mycotoxin risks in the feed industry

Smarter management of mycotoxin risks in the feed industry

For feed millers, the threat of mycotoxin contamination of crops must be handled already at raw material intake. Learn how fast, accurate and reliable results can help you make better decisions about when to safely accept raw material, segregate products or take calculated risks when adding toxin binders.

Slides

Let anyone test when needed

Make fast decisions about raw material intake, segregation and process adjustments such as the use of additive binders and ensure compliance with legal requirements. Avoid human error with a fully automated process that requires no special skills. Just place the grinded sample and push start.

Make better decisions

MycoFoss™ enables testing of up to 6 mycotoxins in one analysis. Increase your knowledge about mycotoxin contamination levels and make more informed decisions with accurate data.

Sample in MycoFoss

Multiplex test for DON and ZEA in barley

A new multiplex test option for the FOSS MycoFoss™ analyzer makes control of barley for feed and malting simpler and more consistent. MycoFoss Global Product Manager, Daniela Vega Sampedro, explains how the long- awaited test was created using rock-solid data references. Learn more

Technical specification

Consumables Ready-to-use sample cups, extraction liquid, assays, standards and flow cytometer cleaning liquid for 100 analyses
Sample volume 4-6 g
Sample preparation The milled sample must have a particle size in which 95 % of the material passes through a sieve of 0.845 mm, 20-mesh or 850 microns. The milled sample should have a flour like texture for proper extraction of mycotoxins
Analysis time 7:55 minutes from inserting sample cup to result
Ambient temperature 7-33 °C

 

 

Dimensions (W x D x H) 700 x 485 x 435 mm. Keep minimum 20 cm between instrument and walls.
Weight 37 kg
Power supply 100-240 VAC ±10 %, 50/60 Hz
Operation environment Indoor facilities. Place the instrument on a stable bench. Avoid accumulation of dust.
Drain Waste container**

 

**) Dispose and label of waste from the instrument according to local legislation.

 

Optimize your instrument performance
Digital services

Streamline your operations and improve your profitability with our FOSS IQX™ digital solutions.

Service agreement

With a FOSS service agreement you will secure operational uptime, regulatory compliance, and analytical accuracy.

Measuring ranges

Singleplex solutions - ppb (µg/kg)
Mycotoxin Corn Wheat Barley Corn DDGS
Aflatoxins - AFLA 2 - 300 - - -
Deoxynivalenol - DON 200 - 30,000 200 - 30,000 200 - 30,000 -
Multiplex solutions - ppb (µg/kg)
Mycotoxin Corn Wheat Barley Corn DDGS
Aflatoxins - AFLA 2 - 100 4 - 100 - 20 - 100
Deoxynivalenol - DON 200 - 5,000 200 - 5,000 300 - 5,000 500 - 5,000
Zearalenone - ZEA 30 - 500 30 - 500 30 - 500 50 - 500
Fumonisins - FUM 500 - 5,000 500 - 5,000 - 500 - 5,000
Ochratoxin A - OTA 5 - 100 5 - 100 20 - 100* -
Trichothecene - T2+HT2 50 - 500 50 - 500 50 - 500* -

 

*) Experimental mode; accuracy of +/-30% stdev across whole measurement range not verified.