NIRS™ DS3 meat by-product analyser

NIRS™ DS3 meat by-product analyser

The NIRS™ DS3 provides NIR analysis of meat by-products with exceptional accuracy. Designed for use in the laboratory, the NIRS DS3 is ideal for routine control.

Value

Value

Analyse samples from rendering production or other meat by-products with one simple analysis in less than a minute and increase your sampling frequency without extra cost.

Type sample

Type sample

Dry samples such as meat and bone meal. Liquid samples such as fats for petfood or energy production. Plant-based products.

Parameters

Parameters

Animal by-products: Fat, moisture, protein, ash. Fats and Oils: FFA, IV, PV, moisture.

Slides

Automated sample transfer

Automated sample transfer reduces time spent on transferring data to and from management systems

Smart Start

The Smart Start function ensures that your NIRS DS3 is ready to measure without having to wait for diagnostics. Just start the shift and start measuring.

FossManager™

FossManager™ software allows instruments to be connected in a network making it easier, quicker and more reliable to manage individual or multiple instruments from a desktop located anywhere.

Ready-to-use calibrations or tools to develop your own

FOSS global ANN calibrations are built on thousands of samples from a large variety of raw materials. Alternatively, develop your own calibrations with the easy to use FossCalibrator™ tool.

Versatile performance

Advanced and versatile solution for testing a range of parameters on ground or liquid material with no or very limited sample preparation.

Meat process image

Improve process and quality control in rendering production with NIR

Learn how rapid meat and bone meal analysis brings value throughout the rendering process. This step-by-step animation illustrates where and how in-line NIR analysis, supported by a NIRS DS3 for validation of results, delivers significant benefits in meat and bone meal production. Learn more

Technical specification

Products measured and product presentation
Feature Specification
Dimensions (w x d x h) 375 x 490 x 300 mm
Weight
27 kg 
Degree of protection IP 65
Measurement mode  Reflectance or transflectance (for liquids) 
Wavelength range
400 - 2500 nm 
Detector
Silicon (400 - 1100 nm), Lead Sulphide, PbS (1100 - 2500 nm) 
Optical bandwidth 8.75 ± 0.1 nm 
Data resolution  0.5 nm 
Number of data points

4200

Number of subsamples Default: 7 for small cup; 8 for large cup
Self test 12 minutes (variable)
Analysis time <1 minute for 32 scans (8 subsamples, 4 scans per subsample)
Network connections
FossManager™
 
Installation requirements  
Voltage supply
100-240 V AC*, frequency 50-60 Hz, Class 1, protective earth
Ambient temperature  5 - 40°C
Storage temperature
-20 to 70°C
Ambient humidity 
< 93% RH
Mechanical environment
Stationary during use
EMC environment
Laboratory use, industry requirements
*Mains supply voltage fluctuations not exceeding ±10% of the rated voltage.
   