Soxtec™ 8000

Soxtec™ 8000

The Soxtec™ 8000 extraction unit is a fully automated system for fast and safe extraction of soluble matter from a wide range of matrices. It performs the four extraction steps boiling, rinsing, solvent recovery and auto-shut down, fully unattended. Just load, start and walk away.

Value

Value

First fully automated Soxhlet system. Improving throughput at a lower cost. Reducing risk in fat analysis. Batch handling tools for easy operation.

Type sample

Type sample

Feed, feed ingredients, forage, pet food, food, food ingredients, grain, cereal, oil seeds, DDGS, meat, meat products, fish, fish products, dairy products, environmental products, (e.g. soil, sludge, waste water) and other industry products (paper pulp, textiles, rubber, tobacco, fertilizer, detergents, plastic and more).

Parameters

Parameters

Crude fat, total fat and extractable matter.

Webinar: How to save time and money with automated fat
                analysis

Webinar: How to save time and money with automated fat analysis

Slides

Flexible operation

The Soxtec™ 8000 total fat solution helps you respond to the demands of your customers, for example, you can do total or crude fat analysis or both. It’s easy to increase capacity according to customer requirements with up to 12 position hydrolysis and six position extraction that is expandable to 12, all with individual hotplate control.

Simplified process

This unique patented Hydrocap™ filter contains the sample from the initial weighing, through hydrolysis to final extraction; avoiding sample transfer errors. It is a fully automated system that performs automated acid hydrolysis and Randall improved Soxhlet extraction in one integrated process.

Secure solvent dispensing & recovery

Accessories

Soxtec™ systems has an extensive range of accessories with glass and aluminium cups and thimbles in different sizes to fit individual laboratory needs and applications.

Soxtec 8000 button

Soxtec™ 8000 integrated total fat analysis

The total fat solution allows you to perform total fat analysis with one integrated process.

There are three main elements to the solution: the Hydrotec™ 8000 hydrolysis unit, the Soxtec™ 8000 extraction unit and the Hydrocap™ that works in both. The extraction unit has an external controller and a broad range of accessories.

Learn more about Hydrotec™ 8000.
Learn more

Soxhlet fat analysis revisited

It’s the first ever fully automated solution for total fat analysis with the Soxhlet method, but how does it work and how does it make life easier in the lab? Read article

Technical specification

Performance data at 230 V
Extraction time
Typically 45 – 60 min
Solvent volume 40 – 110 ml depending on cup type
Sample size 0.5 – 3 g depending on sample type (for crude fat)
0.5 – 2 g (for total fat)
Capacity/batch  6/12 positions
Capacity/day
42/84 samples
Temperature range 0 – 285 °C

Measuring range

0.1 – 100 %

Reproducibility

1% rel. Or better (5 – 100% fat)

Solvent Recovery

Typically 80 %

Accuracy According to officially approved methods
Installation requirements
Power supply 

200 – 240 V 50 – 60 Hz

Power consumption

1500 W (120 VAC version)
1800 W (230 VAC version)

Water consumption

2 l/min < 25°C minimum flow

Weight net

Extraction Unit

35 kg

Control unit

4 kg

Dimensions (W×D×H)

640 × 350 × 630 mm 280 × 200 × 190 mm

A Soxtec™ to suit every need

The Soxtec™ series consist of various levels of automation from manual to fully automated solutions:


Soxtec™ 8000
ST 255 Soxtec™ ST 243 Soxtec™
Levels of automation Fully automated Semi automated Manual
Positions 6 - 12 6 6
Capacity/day 42/86 samples 30 - 36 samples 30 - 36 samples
Solvent volume 40 - 110 ml

70 - 90 ml

40 - 50 ml
Thimble diameter 26, 33 mm 26, 33 mm 26 mm
Compatible with hydrolysis unit Hydrotec™ 8000  Hydrotec™ 8000